  • #clinical trial
  • #atrial pacing
  • A two-lead CRT-DX system without atrial pacing is non-inferior to conventional three-lead CRT-D in patients without sinus node dysfunction.
  • The primary composite endpoint (all-cause mortality, cardiovascular hospitalization, lead complications) showed no significant difference between groups (HR 0.82, 95% CI 0.54-1.25).
  • Fewer atrial lead-related complications occurred with CRT-DX (1.3% vs 4.2%).
  • Echocardiographic reverse remodeling and 6-minute walk test distance were similar between groups.
  • Only one CRT-DX patient required a standard atrial lead during follow-up.