  • Comparison of pulsed field ablation (PFA) and radiofrequency ablation (RFA) in concomitant pulmonary vein isolation (PVI) and left atrial appendage occlusion (LAAO).
  • PFA associated with lower peridevice leak (PDL) rates (5.1%) compared to RFA (23%) at 45-day follow-up.
  • Mean procedure duration significantly shorter with PFA + LAAO (67.1 min) vs. RFA + LAAO (99.6 min).
  • No large PDL (>5 mm) observed; small PDL (≤5 mm) more common in RFA group.
  • Safety outcomes comparable across groups with rare major complications and device-related thrombosis.
  • Study suggests PFA + LAAO improves procedural efficiency and reduces PDL without compromising safety.